-Anwendbare Spritzengröße
2ml, 5ml, 10ml, 20ml, 30ml, 50/60ml
-Rate Bereich
2ml:0.01~60ml, 5ml:0.01~150ml, 10ml:0.01~300ml,
20ml:0,01~600ml, 30ml:0,01~900ml, 50/60ml:0,01~2200ml
-Rate Schritt
0.01ml/h
-BOLUS Rate & Volumen
2ml:0,01~48ml/h 0,1~0,5ml
5ml:0,01~120ml/h 0,1~1ml
10ml:0,01~240ml/h 0,1~5ml
20ml:0,01~480ml/h 0,1~10ml
30ml:0,01~720ml/h 0,1~10ml
50/60ml:0,01~1500ml/h 0,1~10ml
-BOLUS-Rate Schritt
0.01ml/h
-FF(Fast Forward) Rate
2ml:0.01~60ml, 5ml:0.01~150ml, 10ml:0.01~300ml,
20ml:0.01~600ml, 30ml:0.01~900ml, 50/60ml:0.01~2200ml
-FF vol.
0 ~maximales Volumen der Spritze
-KVO-Rate
0.01~5,0ml/h (0,01ml-Schritt)
-KVO-Vol.
2ml:0~0.2ml, 5ml:0~0.2ml, 10ml:0~0.2ml,
20ml:0~0.5ml, 30ml:0~0.5ml, 50/60ml:0~1ml (0.01ml Schritt)
-VTBI (zu infundierendes Volumen)
0.01~9999.99ml, (0.01ml Schritt)
-Kumulative Inf.vol.
0.01~9999,99ml, (0,01ml Schritt)
-Ratengenauigkeit innerhalb ±2%
Druckeinheiten MPa, kPa, mmHg, inH2O, PSI, mbar
-Okklusionsdruck
Stufe 1-13 (10,20,30,40,50,60,70,80,90,100,110,120,130)kPa
-Informationen zu Alarmen
infusion beendet, kurz vor Ende der Infusion, VTBI beendet, System-Standby beendet, Infusion blockiert, Spritze fällt ab, Spritze ist nicht richtig installiert, vergessene Bedienung, schwache Batterie, leere Batterie, abnormaler Betriebszustand, Gerätefehler, Stromausfall usw.
-Spritzenhöchstwert (unter einer einzigen Fehlerbedingung)
1mL
-Stromquelle, Stromverbrauch, Sicherung
100~240VAC, 50/60Hz, 25VA, T2A/250VAC
Interner wiederaufladbarer Li-Akku 11.1V, Kapazität≥2600mAh, 8 Stunden Backup-Zeit bei 5ml/h
Externes DC-Netzteil, Eingangsspannung 15V, Eingangsstrom 1,2-1A.
-Klassifizierung der Ausrüstung
Klasse I, Typ CF, Dauerbetrieb
-IP-Klasse
SCHUTZART 24
-Abmessungen & Nettogewicht
301mm(L)×183mm(B)×82mm(H) & 2,0kg
---